AED7000 automatic external defibrillator
AED7000 automatic external defibrillator Three-step defibrillation process Two button operation A wide range of voice and visual prompts to the operator Two-stage energy output
External lock protection prevents unintentional defibrillation Defibrillation characteristics Defibrillation output waveform: biphasic truncated exponential waveform
Energy sequence: 150, 150, 200J
Charging time: When the battery is used less than 15 times, the maximum charging energy
is less than 10 seconds Analysis time: 9 seconds
In the case of a new battery, the maximum time from the beginning of the heart rate
analysis to the preparation of discharge: <25 seconds After six discharges, the maximum time from the start of heart rate analysis to the
preparation of discharge: <35 seconds Voice prompt: 20 prompt tones Visibility reminder: LED reminder Control: two buttons, on/off and discharge Accuracy of discharge energy: within the range of 20 to 200Ω, the discharge accuracy is ±15%
Maximum voltage: 1100±50V
Defibrillation impedance range: 20 to 200Ω