Performance | |
Method | HIGH pressure ion exchange liquid chromatography (HPLC) |
Instrument | Fully-Auto Glycated Hemoglobin HbA1c Analyzer |
Sampling system | Automatic identification system for Primary Tube & sample cup, Primary tube with cap puncture automatically |
Throughput | ≤3mins per sample |
Sample Tray | 15 samples position, CAL1(QC1) & CAL2(QC2)position, ST |
Sample Type | Venous Blood, Finger Peripheral Blood, or hemolyse blood |
Sample Volume | 5ul whole blood, or lyse blood (ratio 5ul:1000ul) |
Test Range | HbA1c: 3-18%; 14-191mmol/mol, HbF:0.3-5% |
Test Item | HbA1c HbF & β-Thalassemia |
Heat system | Temperature range for column 5-40℃, can keep the constant temperature, the fluctuation £±0.2℃ |
Light system | 415nm LED light, long life span, the fluctuation is £2mv per hour |
Printer | 57nm Thermal printer, printing test curve and report |
USB Port | Support LIS system, software upgrade online, external barcode scanner optional |
Barcode | External barcode scanner optional |
Display | 8 inch LCD colorful touch screen, 162mm´122mm, display real-time testing curve |
Data Storage | more than 10,000 test reports storage (including Curve),Inquiry freely |
Language | English, Spanish, Russians, Portuguese, Japanese…… |
Report | NGSP, IFCC result HbA1a, HbF, HbA1c, HbA0,HbA2 variant sample report. Optional PC software for regular report format |
Work Condition | Temperature 5℃~30℃, Relative humidity 30%~75% |
Power supply | 110V-220V, 50-60Hz |
Dimension | 460mm×445mm×390mm |
Weight | 28KGS |
Within-Run(n=20) | Inter-Run(n=20) | |||
Low | High | Low | High | |
Mean | 5.56 | 10.24 | 5.64 | 10.18 |
SD | 0.08 | 0.09 | 0.12 | 0.15 |
CV(%) | 1.43 | 0.87 | 2.13 | 1.47 |