Main technical parameter 6.1. Power supply: ~ 220V, 50/60 Hz 6.2. Power input: 1100VA 6.3. Normal working condition: a. Environment temperature: 18℃~30℃; b. Relative humidity: 30%~75%; c. Atmospheric pressure: 700~1060hPa; d. Airvelocity: <0.3m/s 6.4. Temperature control range: 32℃~38℃ 6.5. Precision of temperature control: ≤0.5℃ 6.6. Functional alarm: 6.6.1. Over-temp: 39℃ The secondary over temp to shut off power: 40℃(cut off the heating power, alarm with sound and light) 6.6.2. Deviation alarm: When the temp goes steady and the deviation reaches ±1℃(+1℃, the heating power will be shut off), alarm with sound and light. 6.6.3. Sensor alarm: When the skin temp sensor is in short circuit or open circuit state, shut off the heating power, alarm with sound and light; the sensor overruns the radiation area and temp deviation reaches to –1℃ for about one minute, the machine will alarm with sound and light automatically and keep certain power to heat up. 6.6.4. Power failure alarm: When the power breaks off, alarm with sound and light. 6.7. Bed temp uniformity: ≤2℃ 6.8.Skin temp sensor precision: ≤0.3℃ 6.9.Increasing temp time: 45’ 6.10. Transportation and storage a.Environment temp: -40℃~+55℃ b.Relative humidity range: ≤95% c. Atmospheric pressure range: 500~1060hPa