YSAV90A hospital ICU oxygen respirator ventilator

Item No.: YSAV90A
Gas requirement: Medical oxygen and nitrous oxide with a pressure ranging from 280~600kpa
safety valve <12.5  kPa
Respiratory Rate 1 ~ 99 bpm
gas concentration output 21% ~ 100%  adjustable
Maximum Minute ventilation ≥18L / min
modes of ventilation
  • BrandYsenmed
  • Model numberYSAV90A
  • Warrantyone year
  • Customizedyes
  • Lead time15 working days
  • Packagecarton packing
  • Description
    YSAV90A  hospital ICU oxygen respirator ventilator
     
    TECHNIQUES SPECIFICATIONS
    Physical specifications
    Display  10.4'' LCD Display
    Mode: pneumatically powered and electrically controlled system
    Trolley: Fitted with 4 nos anti-static rubber castors; two of which are lockable for braking and easy maneuverability with foot operated brake provisions
    Technical Specifications:
    Gas requirement: Medical oxygen and nitrous oxide with a pressure ranging from 280~600kpa
    safety valve <12.5  kPa
    Respiratory Rate 1 ~ 99 bpm
    gas concentration output 21% ~ 100%  adjustable
    Maximum Minute ventilation 18L / min
    modes of ventilation A/C, SIPPVIPPVSIMVPSIMVCPAPPCVPEEPSIGNSPONT, MANUAL
    Ventilator compliance 30 Ml/kPa
    PEEP: 0 ~ 1.0 kPa
    Ptr: -10 ~ 10 hPa
    IMV frequency: 1~20 bpm
    I/E Ratio: 4:1~1:4, Has the inverse ratio ventilation
    Tidal Volume Setting Range:  50~1500ML   Display Range: 0~2000ml ;  resolution ratio: 10ml
    Inspiratory Plateau: 0~1s
    Maximum Safety  Pressure ≤ 6.0kPa
    Sigh: one deep breath per 80 controlled respirations, the inspiration time is 1.5 times of the setting point
    Pressure limit range: 1.0~ 6.0kPa
    Airway pressure alarm: Audible and visual and with yellow and red color indicating Lower: 0.1 kPa ~ 6.0 kPa;   Upper: 0 ~ 5.0  kPa
    ±0.2 kPa
    Ventilation Volume alarm: ≥ 18L/ MIN
    Power Supply Alarm Ac/dc power supply are after failing to send out alarm immediately
    Alarm Time: keep >120s
    The airway pressure continues to be higher than 15 hPa ±1 hPa for 15s±1s, then the machine will raise an audible alarm, the pressure will be displayed in red and the continuous high pressure red alarm wording is displayed on the screen of the anesthetic respirator.







    a