NIBP
Method: Oscillometry
Measure mode: Manual, Auto, STAT
Measure interval in auto mode:
1~5/10/15/30/60/90/120/180/240/480min
Measure period in STAT mode: 5min
Pulse rate range: 40 ~ 240bpm
Measure & alarm range: Adult
SYS 40 ~ 280mmHg
DIA 10 ~ 220mmHg
MEAN 20 ~ 240mmHg
Pediatric
SYS 40 ~ 220mmHg
DIA 10 ~ 160mmHg
MEAN 20 ~ 170mmHg
Neonatal
SYS 40 ~ 135mmHg
DIA 10 ~ 100mmHg
MEAN 20 ~ 110mmHg
Static pressure measurement
Range: 0 ~ 300mmHg
Accuracy: ± 3mmHg
Resolution: 1mmHg Accuracy:
Max. Mean Error: ± 5mmHg
Max. Standard Deviation: 8mmHg
Overpressure protection:
Adult 300mmHg
Pediatric 240mmHg
Neonatal 150mmHg
SPO2
Measurement range: 0 ~ 100%
Resolution: 1%
Accuracy: 70% ~ 100% ± 2%
40% ~ 69% ± 3%
Environment
Operating environment :
Temperature: 0 ~ 40 ℃
Humidity: 15% ~ 80%
Pressure: 86.0 kPa ~ 106.0kPa
Transport and Storage environment :
Temperature: -20 ~ 55 ℃
Humidity: 10% ~ 90%
Pressure: 86.0 kPa ~ 106.0kPa
YSPM80C MULTI-PARAMETER PATIENT MONITOR
Model NO.: YK-8000C
Quality Standards and Classification
CE, ISO13485
SFDA: Class Ⅱb
Anti-electroshock degree:
ClassⅠequipment
(internal power supply)
TEMP/SpO 2 /NIBP:
YSPM90D 15 inch screen medical patient monitor
1.15" high resolution color TFT LCD display;
2.13 Type Arrhythmic Analysis, Multi_Lead ECG Waveforms Display in Phase, Real time S_T segment analysis, pacemaker detection Drug calculation and titratiotable;
3. Efficient resistance to interference of defibrillator and electrosurgical cautery;
4. SPO2 can testing for 0.1% Weak;
5. RA-LL impedance Respiration;
6. Trend Coexist Display;
7. OxyCRG DyNamic View Display;
8. Bed to Bed view Display;
9. Networking capacity ;
10. Options of printing, IBP, VGA and EtCO2;
11. UP to 4 hours working capacity of built_in rechargeable battery;
12.15" high resolution color TFT LCD display;
13. Large volume of tabular and graphic trends information storage and easy to recall;
14. Anti-ESU, anti- defibrillator;
15. Capture dynamic waveforms.
Technical specification
Specifications:
ECG
Lead Mode: 5 Leads ( R, L, F, N, C or RA, LA, LL, RL,V)
Lead selection: I, II, III, avR, avL, avF, V,
Waveform: 2 ch
Lead mode: 3 Leads ( R, L, F or RA, LA, LL)
Lead selection: I, II, III,
Waveform: 1 ch
Gain: ′2.5mm/mV, ′5.0mm/mV, ′10mm/mV, ′20mm/mV, auto
HR and Alarm
Range
Adult: 15 ~ 300 bpm
Neo/Ped: 15 ~ 350 bpm
Accuracy : ± 1% or ± 1bpm,which great
Resolution: 1 bpm
Sensitivity: > 200 (uV P-P)
Differential Input Impedance: > 5 MΩ
CMRR
Monitor: > 105 dB
Operation: > 105 dB
Diagnosis: > 85 dB
Electrode offset potential: ±300mV
Leakage Current: < 10 uA
Baseline Recovery: < 3 S After Defi.
ECG Signal Range: ±8 m V ( Vp-p )
Bandwidth
Surgery: 1 ~ 15 Hz
Monitor: 0.5 ~ 35 Hz
Diagnostic: 0.05 ~ 100 Hz
Calibration Signal: 1 (mV p-p), Accuracy:±5%
ST Segment Monitoring Range: Measure and Alarm -2.0 ~ +2.0 mV
ARR Detecting
Type: ASYSTOLE, VFIB/VTAC, COUPLET, BIGEMINY, TRIGEMINY, R ON T, VT>2, PVC,
TACHY, BRADY, MISSED BEATS, PNP, PNC
Alarm: Available
Review: Available
RESPARATION
Method: Impedance between R-F(RA-LL)
Differential Input Impedance: >2.5 MΩ
Measuring Impedance Range: 0.3~5.0Ω
Base line Impedance Range: 0 – 2.5 KΩ
Bandwidth: 0.3 ~ 2.5 Hz
Resp. Rate
Measuring and Alarm Range
Adult: 0 ~ 120 rpm
Neo/Ped: 0 ~ 150 rpm
Resolution: 1 rpm
Accuracy : ±2 rpm
Apean Alarm: 10 ~ 40 S
NIBP
Method: Oscillometric
Mode: Manual, Auto, STAT
Measuring Interval in AUTO Mode: 1, 2, 3, 4, 5, 10, 15, 30, 60, 90, 120, 180, 240,480 (Min)
Measuring Period in STAT Mode: 5 Min
Pulse Rate Range: 40 ~ 240 bpm
Alarm Type: SYS, DIA, MEAN
Measuring and alarm range
Adult Mode
SYS: 40 ~ 270 mmHg
DIA: 10 ~ 215 mmHg
MEAN: 20 ~ 235 mmHg