Feature
Portable and Adjustable:
12.1" screen, adjustable from 0 to 90 degrees;
Data Storage:
Stores 24 hours of CTG data with power-off memory;
Display:
Adjustable background, shows waveform and data together;
Printing:
Utilizes unique thermal printing for wide-range, high-speed printing(25mm/s);
Fetal Monitoring:
Tracks FHR, TOCO, FM, and fetal activities, with automatic and manual fetal movement count display modes,
Power Efficiency:
Low-power consumption design with standby mode and rechargeable battery for 24-hour monitoring;
User Management:
Allows auto/manual patient ID generation, easy user management, and data searching;
Printing Options:
Supports real-time and timed printing with event remark functionality;
Electrosurgical Compatibility:
Features patented technology to maintain ECG waveform and HR stability during procedures;
NIBP Accuracy:
Uses Goertzel arithmetic for fast, accurate NIBP readings, including venipuncture support;
Alarm System:
Equipped with audible and visual alarms, adjustable limits and tones, with sensor off, abnormal alarm, remark, and alarm review functions
Data Transmission:
Offers USB output for easy data transfer.
Specifications
FHR
Ultrasound frequency: 1MHZ±1KHZ
Ultrasound input intensity: <1mW/cm2
Range: 30~240bpm
Accuracy: ±1bpm
TOCO
Measurement range: 0~100
Resolution: 1
Pressure tolerance: ≤±8%
Zero-mode: Manual
Fetal Movement
Count mode: Manual,automatic and automatic fetal movement waveform
Printing paper: Foldable Z-type thermal paper
Width: 152mm
Sweep speed: 1cm/min,2cm/min,3cm/min,25mm/sec,quick recall printing
ECG
Input: 3/5 wires ECG cable
Lead section: I,ll, lll, aVR, aVL,aVF,V
Gain selection: x1/4,x1/2,x1,x2,x4 and automatic
Sweep speed: 3.125,6.25,12.5,25,50mm/s
Heart rate range: 15~350bpm
Calibration: ±1mv
Accuracy: ±1bpm or ±1%(select the bigger data)
SpO2
Display type: Waveform, Data
Measurement range: 0~100%
Accuracy: ±3%(between 70%~100%)
SpO2 resolution: 1%
Pulse rate range: 25~250bpm
Accuracy: ±1bpm or±2% (select the bigger data)
PR Resolution: 1bpm
PI
Measurement range: 0.00~20.0%
Accuracy: Not defined
Resolution: 0.00~9.99%:0.01%
10.0~20.0%:0.1%
NIBP
Philosophy: Oscillometric method
Measurement parameter: Systolic, Diastolic, Mean
Measurement method: Manual, Auto, Continuous
Measuring range: SYS:30-270mmHg
DIA:10-200mmHg
MAP:20-220mmHg
Resolution: 1mmHg
Accuracy: Max average error: ±5mmHg
Max standard deviation:8mmHg
2-Temperature (Rectal & Surface)
Number of channel: 2 channel
Measurement range: 0°C~50°C
Accuracy: +0.2°C
Resolution: 0.1°C
Display: T1,T2,∆T
Unit: °C/°F selection
2-Respiration(Impedance & Nasal Tube)
Measurement range: 0~120rpm
Accuracy: ±1rpm or ±5%
(select the bigger data)
Resolution: 1rpm
Configuration
YSFM-G6B Standard configuration
FHR, TOCO,FM, Built-in printer and battery
YSFM-G6B Plus Standard configuration
FHR, TOCO, FM, Built-in printer ECG, NIBP, SPO2,2RESP,2TEMP, HR/PR, Rechargeable battery
Optional configuration
Twin FHR, Wireless Probe, Fetal Wake Up, Fisher Auto Fetal-Analysis, Touch Screen, Wireless Network,12-Volt Power Supply, High Capacity Battery, Wall Mount, Moving Cart