| Model | YSTE310K | |
| Overall Performance |
Throughput | 300Tests/hour, ISE(Optional) |
| Analysis Method | End-point, Fixed-time, Rate(Kinetic), Turbidimetry | |
| Certificates | CE, FDA, ISO9001, ISO14001,1SO13485 | |
| System | Optional open or close system, continuous working for 24 hours | |
| Sample & Reagent System |
Sample Positions | 49 sample positions |
| Reagent Position | 56 reagent positions | |
| Sample Volume | 2-70μL | |
| Reagent Volume | 20-350μL | |
| Probe | With anti-collision function, liquid level detection function | |
| Probe Washing | Automatic washing interior and exterior | |
| Reagent Cooling | 2~12°C for 24 hours | |
| Reaction System | Temperature Control | 37±0.2°C, real-time monitoring |
| Cuvettes | 120 reusable cuvettes | |
| Mixer Probe | Independent stirring, automatic frequency conversion | |
| Washing | Automatic cuvettes washing | |
| Optical System | Light Source | Halogen lamp, water-cooling |
| Spectrophotometry | Post-spectral spectrophotometry | |
| Wavelength | 340~700nm | |
| Absorbance | 0~3.5Abs | |
| Calibration & QC | Calibration | Linear:K factor, 1-point,2-point and multipoint linear Non-Linear: Spline, Polygon, index, Ogarithm, Logit-4P, Logit-5P |
| Quality Control | Real-time QC, Westgard multi rule, Cumulative Sum Check, Twin Plot(2D) | |
| Data Management | Software | Windows 7/8/10, 32or 64bit |
| LIS System | Bi-direction, support HL7 protocol | |
| Interface | LAN port access | |
| Printer | External, multiple reporting mode available | |
| Working Conditions | Power supply | AC220V±10%,50/60Hz(Standard),110V±10% 60Hz(Optional),350VA |
| Temperature | 15~30°C (+2°C) | |
| Water Consumption | Deionized water<5L/h | |
| Humidity | 40~85% | |
| Size & Weight | External Size (W*D*H) | 873*585*500mm |
| Net Weight | 80kg | |
| Package Size (W*D*H) | 1130*735*1040mm | |
| Gross Weight | 125kg | |