Model | YSZS-DSP5 | YSZS-DSP5A | YSZS-DSP5P |
Applicable Syringe Size | 5/6ml、10/12ml、20ml | 30/35m、50/60ml | 2/3ml、5/6ml、10/12ml 20ml、30/35ml、50/60ml |
Rate Range/FF Rate /Bolus Rate |
5/6ml:0.01~300mlh 20ml:0.01~1200ml/h 50/60ml:0.01~2400ml/h |
10/12ml:0.01~600ml/h 30/35ml:0.01~1800ml/h |
2/3ml:0.01~160ml/h 5/6ml:0.01~400ml/h 10/12ml:0.01~800ml/h 20ml:0.01~1600ml/h 30/35ml:0.01~2400ml/h 50/60ml:0.01~3000ml/h |
WIFI | N/A | YES | Optional |
Pressure Level | 13 Level | 13 Level | 16 Level 14 and above, 50ml/60ml syringe is not applicable |
KVO Rate (manual mode) | 0.01~30ml/h(0.01ml/h step) | ||
Rate accuracy | within ±1.8%(after calibration) | ||
VTBI (volume to be infused) | 0~9999.99ml, 0.01mlstep(value"0" means no limit for VTBI) | ||
Cumulative Injection Volume | 0~9999.99ml, (0.01ml step) | ||
Occlusion Pressure | Level 1-16(10, 20, 30, 40, 50, 60, 70, 80, 90, 100, 110, 120, 130, 140, 150, 200) kPa (level14, 15, 16 can not applicable to 50/60mlsyringes) | ||
Alarms Information | Near infusion end, Infusion finished, VTBI completed, Near VTBI end, extension tubing off(optional), standby task completed, infusion is blocked, syringe falls off, syringe is improperly installed, forgotten operation, low battery, battery empty, equipment failure, system power down, battery disconnected, communication interruption, standby task completed, cascade failure, cascade completed | ||
Dimensions &Weight | 384mm(L)x223mm(W)x231mm(H)/4.6kg |