Main body
|
High strength engineering plastic rack, light,beautiful and corrosion
resistant |
||
Scope of application
|
adult, child
|
|
|
Gas source
|
O2: 0.27~0.55MPa N2O: 0.27~0.55MPa
|
||
Flow meter
|
O2: 0.1~1.0L/min 1.1~10L/min
|
||
N20: 0.1~1.0L/min 1.1~10L/min
|
|||
O2, N2O linkage and N2O Stopper
When using nitrous oxide, oxygen concentration>25%; When the oxygen pressure<0.2Kpa, the flow of nitrous oxide would be cut off. |
|||
Flow rate of rapid oxygen supply
|
25~75L/min
|
||
Low oxygen pressure alarm
|
There will be sound alarm when the oxygen
pressure < 0.2MPa |
||
Vaporizer
|
it has the function of automatic compensation
based on pressure, temperature, and flow rate. The regulation range of evaporator concentration is 0~5 vol%. Among Halothane,Enflurane,Isoflurane and Sevoflurane,two can be chosen for application as required by the customer. And vaporizers that are imported with original packaging are also available. |
||
Respiratory circuit
|
working mode: all-close, semi-close, semi-open
APL: 0.5~7 kPa |
||
Respiratory Bellows
|
bellows for adults, bellows for children
tidal volume range:0~1,500 ml |
||
Ventilator
|
|||
Display mode
|
High-definition 5.7” LCD screen display
|
||
Ventilation mode
|
IPPV, SIPPV, SIMV, PEEP, MANUAL, SIGH
|
||
Ventilation parameters
|
Tidal volume
|
50~1500 ml
|
|
Rate
|
2~99 bpm
|
||
SIMV rate
|
2~20 bpm
|
||
I:E
|
2:1~1:8
|
||
Inspiratory trigger pressure
|
-1.0~2.0 kPa
|
||
PEEP
|
0~2.0 kPa
|
||
Pressure range
|
1.0~6.0 kPa
|
||
SIGH
|
1.5 times the inspiratory time
|
||
Parameters for ventilation
monitoring |
tidal volume, ventilation volume, IPPV rate,
SIMV rate, total respiratory rate, I/E, peak pressure of airway, pressure - time waveform, flow rate - time waveform, PEEP, inspiratory trigger pressure |
||
Monitoring of oxygen
concentration |
(21%~100%)
|
||
Security Alarm System
|
|||
Oxygen Concentration alarm
|
upper limit setting range (50%~100%)
low limit setting range (15%~50%) |
||
Airway pressure alarm
|
upper limit setting range 1.0~6.0 kPa
low limit setting range 0.4~2.0 kPa |
||
Per-minute ventilation volume alarm
|
upper limit setting range 3.0~30L/min
low limit setting range 1.0~10L/min |
||
Sustained high-pressure alarm
|
it will give alarm when stress have consistently been higher than 2.5 kPa
|
||
Suffocation alarm
|
it will give sound and light alarm if there is no tidal volume input
for 15 seconds |
||
Power alarm
|
|||
Power
|
AC 220V±10% 50Hz±1Hz(UPS, Storage battery)
|
||
(Mechanical Arms)
|
|||
(End-expiratory CO2 monitoring)
Items within brackets are optional. |