| No. | Item | Specification |
| 1 | X-ray Generator | |
| 1.1 | Generator Type | High Frequency Inverter 80kHz |
| 1.2 | Input Power | Single Phase 220VAC, 50/60Hz |
| 1.3 | Radiographic Ratings | Large Focal Point 20-35kV/10-510mAs |
| Small Focal Point 20-35kV/10-100mAs | ||
| 1.4 | Power Rating | 6.2kW |
| 2 | X-ray Tube | |
| 2.1 | Model | China Hangzhou LR01 |
| 2.2 | Focal Spot Size | Dual Focus 0.2 / 0.4mm |
| 2.3 | Target Material | Molybdenum (Mo) |
| 2.4 | Port Material | Beryllium (Be) |
| 2.5 | High-speed Anode Drive | 2800rpm |
| 2.6 | Target Angle | 12° |
| 2.7 | Anode Heat Storage | 80kJ(110kHU) |
| 2.8 | Anode Cooling | Air Cooling |
| 2.9 | Filtration | Mo(0.03mm) |
| 3 | Radiographic Stand | |
| 3.1 | U-arm Vertical Movement | 590mm |
| 3.2 | Rotations Deg | +90°~-90° |
| 3.3 | SID | 650mm |
| 3.4 | Compressor | The compressor is automatically released after the exposure is completed |
| 3.5 | Compressor Max. Pressure | 200N |
| 3.6 | Compressor Max. Travel | 156mm |
| 4 | Detector | |
| 4.1 | Detector Material | Amorphous Silicon |
| 4.2 | Effective Coverage Of Detector |
24cm×30cm |
| 4.3 | Pixel Matrix | 2816×3528 |
| 4.4 | Limit Of Spatial Resolution | 6.0Lp/mm |
| 4.5 | DQE Value | @4 Lp/mm:≥35% |
| 4.6 | Dynamic Range | 14bit Digital Output |
| 4.7 | Pixel Size | 85μm |
| 4.8 | Output | Ethernet |
| 4.9 | Working Condition | 10℃-40℃ |
| 4.10 | Storage Environment | -10℃-50℃ |
| 5 | Image Acquisition Workstation Hardware |
|
| 5.1 | CPU | Intel Core Duo 3.0GHz |
| 5.2 | Hardware | ≥1T high Speed Hardware |
| 5.3 | Memory | ≥8G |
| 5.4 | Display Card | ≥2G |
| 5.5 | High Brightness High-contrast LCD |
1920*1080Pixel Resolution |
| 5.6 | Network interface Worklist | DICOM3.0 Transmission |
| 100/1000 Gigabit Ethernet | ||
| 5.7 | Software | Imaging Software Package DMOC V1.0 |
| No. | Configuration | Quantity |
| 1 | Mammography machine host | 1 |
| X-ray tube for breast | 1 | |
| High frequency X-ray generator | 1 | |
| Mammography column | 1 | |
| Electric lifting and rotating U-shaped arm | 1 | |
| Console | 1 | |
| 2 | Foot switch | 1 |
| 3 | Exposure hand brake and connection | 1 |
| 4 | Supporting cables include: | - |
| Dedicated three-wire power cord | 1 | |
| Dedicated ground wire | 1 | |
| Spare fuse | 2 | |
| 5 | Flat panel detector | 1 |
| 6 | Control system of special high-pressure generator for breast | 1 |
| 7 | Breast image acquisition system | 1 |
| DM image acquisition software package | 1 | |
| DMOC image processing module |