Main body | High-strength engineering plastic rack, light, beautiful and corrosion resistant | |
Scope of application | adult, child | |
Gas source | O2: 0.27~0.55MPa N2O: 0.27~0.55MPa | |
Flow meter | O2: 0.1~1.0L/min 1.1~10L/min | |
N20: 0.1~1.0L/min 1.1~10L/min | ||
O2, N2O linkage and N2O Stopper When using nitrous oxide, oxygen concentration>25%; When the oxygen pressure<0.2Kpa, the flow of nitrous oxide would be cut off. |
||
Flow rate of rapid oxygen supply | 25~75L/min | |
Low oxygen pressure alarm |
There will be sound alarm when the oxygen pressure < 0.2MPa |
|
Vaporizer | it has the function of automatic compensation based on pressure, temperature, and flow rate. The regulation range of evaporator concentration is 0~5 vol%. Among Halothane,Enflurane, Isoflurane and Sevoflurane, one can be chosen for application as required by the customer. And vaporizers that are imported with original packaging are also available. |
|
Respiratory circuit |
working mode: all-close, semi-close, semi-open releasing pressure: 0.5~7 kPa |
|
Respiratory Bellows |
bellows for adults, bellows for children tidal volume range:0~1,500 ml |
|
Ventilator | ||
Display mode | High-definition TFT color LCD screen display | |
Ventilation mode |
IPPV, SIPPV, SIMV, PEEP, MANUAL, SIGH, (IMV, VCV, PCV) |
|
Ventilation parameters | Tidal volume | 50~1500 ml (20~1500 ml) |
Rate | 1~99 bpm | |
SIMV rate | 2~20 bpm | |
I:E | 2:1~1:8 (4:1~1:8) | |
Inspiratory trigger pressure | -1.0~2.0Kpa | |
PEEP PressureRange |
0~2.0Kpa 1.0~6.0Kpa |
|
SIGH | 1.5 times the inspiratory time (70-150 adjustable) | |
Inspiratory platform | 0~50% | |
Parameters for ventilation monitoring tidal volume, ventilation volume, IPPV rate, SIMV rate, total respiratory rate, I/E, peak pressure of airway, average pressure, pressure - time waveform, flow rate - time waveform, PEEP, inspiratory trigger pressure, Inspiratory platform (autonomous inspiratory rate, lung compliance, airway resistance,flow-volume oop,pressure-volume loop) | ||
Monitoring of oxygen concentration | 21%~100% | |
Security Alarm System | ||
Oxygen Concentration alarm | upper limit setting range 21%~100% | |
low limit setting range 10%~80% | ||
Airway pressure alarm | upper limit setting range 0.3~6.0 kPa | |
low limit setting range 0.2~5.0 kPa | ||
Per-minute ventilation volume alarm | upper limit setting range 3.0~30L/min | |
low limit setting range 1.0~10L/min | ||
Sustained high-pressure alarm | it will give alarm when stress have consistently been higher than 2.5 kPa | |
Suffocation alarm | it will give sound and light alarm if there is no tidal volume input for 15 seconds |
|
Power alarm | ||
Power | AC 220V±10% 50Hz±1Hz(UPS, Storage battery) | |
(Mechanical Arms)(End-expiratory CO2 monitoring) | ||
Items within brackets are optional. |