| YSTE240 Automatic Chemistry Analyzer | ||||||||
| Specifications | ||||||||
| Analyze System | ||||||||
| Throughput | Maximun 240T/h | |||||||
| Theory | Absorption spectronmetry,Transmission turbidimetry | |||||||
| Methodology | End-point,Fixed-time,Kinetic,Monochromatic/bichromatic | |||||||
| Test according set up,random access | ||||||||
| Reaction plate | Contain 81 reaction cuvettes | |||||||
| Maximun reaction time | 15 minutes | |||||||
| Minimum reaction volume | 150 ul | |||||||
| Material of cuvettes | Special UV-transparent optical glass | |||||||
| One-off cuvettes,can update to automatically cuvettes washing | ||||||||
| Optical System | ||||||||
| Light Source | Halogen-tungsten lamp | |||||||
| Photometer | Reversed optics | |||||||
| Wavelength | 340nm-900nm wavelength option | |||||||
| Absorption | 0-4 Abs | |||||||
| Reagent/Sample Handling | ||||||||
| Sample volume | 2-100 ul,step by 0.1 ul | |||||||
| Sample plate | Contain60 sample positions | |||||||
| Reagent/Sample Probe | Automatically checked,patent technology liquid level detection,collision protection Probe cleaning:Automatic washing for both inside and outside | |||||||
| Samples cup | Support one-off blood cuvettes、micro cups、plastic cuvettes | |||||||
| Reagent volume | 20-500 ul,step by 1 ul | |||||||
| Reagent plate | Contain 20 reagent positions | |||||||
| Sample/reagent probe | Collision protection liquid level detection;Internal and external probe washing;Probe depth adjustment automatically | |||||||
| Reagent bottle | 15ml & 30ml; Dead volume less than 0.5 ml | |||||||
| Temperature Control | ||||||||
| Maintenace free heater elements | ||||||||
| Temperature maintainance within 3 minutes | ||||||||
| Temperature of reagent compartment | 4-12℃ | |||||||
| Reaction temperature | 37±0.1℃ | |||||||
| Temperature fluctuation | ±0.1℃ | |||||||
| Operation System | ||||||||
| Software | WindowsXP、2000 | |||||||
| Database | SQL Server | |||||||
| Data Storage | Automatic store & copy;No time and volumn limit;Lifetime preservation | |||||||
| Print format | Can be set up,offer several standard format | |||||||
| Calibration mode | Linear curce type:Single-point linear,two-point linear and multi-point linear none linear;Curve type:Logistic-Log 4P,Logistic-Log5p,Spline,index automatically use best calibration curve and standard | |||||||
| QC Methods | Real time,within day & between day | |||||||
| Quality Control Diagram | L-J、cumulative、twin plot | |||||||
| Control software | Random set up,default westgard multi-rule | |||||||
| Status display | Real time monitoring of reagent residual volume Running status of reagent tray,samples tray | |||||||
| Adjustment | Dynamic and real time display of running status;Intelligent carry over settings to adjust test sequence;Original reaction data record;Detail profile of alert messages | |||||||
| Setting | Operation permission setting,Item setting,cross polution setting,cuvette blank setting | |||||||
| Others | ||||||||
| Water consumption | Less than 1 L per hour | |||||||
| Power Supply | AC 220V±22V 50Hz±1Hz | |||||||
| Frequency | Maximun 200VA | |||||||
| Dimension | 720mm×581mm×545mm | |||||||
| Weight | 60kg | |||||||
| Accessory | Software | |||||||